Free Access

Table I.

Instrumental and operating conditions for FAAS.

Element Wavelength (nm) Lamp current (mA) Slit width (nm) Background correction Flame chemistrya
Na 589.0 6 0.2 Air/C2H2 – stoichiometric
K 766.5 7 0.5 Air/C2H2 – stoichiometric
Ca 422.6 5 0.5 N2O/C2H2 – reducing
Mg 285.2 3 0.5 Air/C2H2 – stoichiometric
Zn 213.9 8 0.2 D2 lamp Air/C2H2 – stoichiometric
Fe 248.3 10 0.2 D2 lamp Air/C2H2 – stoichiometric
Mn 279.5 10 0.2 Air/C2H2 – stoichiometric

Without background correction.

a

Burner position: aligned with optical axis.