Free Access

Table III.

Performance characteristics of calibration functions.

Element Concentration range (mg·L−1) Absorbance range TV Vxo CLOD (mg·L−1) CLOQ (mg·L−1)
Na 0.02–0.3 0.02–0.21 12.2 0.03 0.008 0.02
K 0.02–0.5 0.02–0.20 7.2 0.02 0.008 0.02
Ca 0.02–4.0 0.03–0.25 2.8 0.03 0.07 0.2
Mg 0.02–0.2 0.08–0.28 34.1 0.02 0.008 0.02
Zn 0.02–0.4 0.02–0.18 0.1 0.03 0.007 0.02
Fe 0.1–1.0 0.02–0.11 1.2 0.05 0.03 0.10
Mn 0.025–0.50 0.01–0.11 1.1 0.01 0.001 0.004
P 0.05–1.0 0.06–0.86 278 0.11 0.01 0.03
Element
Concentration range (μg·L−1)
Absorbance range
TV
Vxo
CLOD (μg·L−1)
CLOQ (μg·L−1)
Cr 0.25–2.5 0.02–0.12 −2.8 0.07 0.1 0.3
Mo 1–10 0.02–0.12 0.2 0.07 0.4 1.0
Ni 1–10 0.02–0.07 1.7 0.04 0.3 1.0
Cu 1–20 0.02–0.30 −2.4 0.04 0.37 1.0
Cd 0.1–1.5 0.02–0.22 −3.0 0.05 0.04 0.1
Co 0.5–7.5 0.02–0.18 −2.8 0.08 0.2 0.6
Pb 1–7.5 0.02–0.10 −3.0 0.10 0.5 1.0

TV, testing value for linearity evaluation ; Vxo, relative variation coefficient [16, 17]; N, number of calibration data pairs; CLOD, limit of detection and CLOQ, limit of quantification (both expressed as concentration of the elements in the milk digestion solutions).